IO-Link Compatible Non-Contact ThermometerTI-S Series

DOWNLOAD

High-Speed & High-Accuracy
Non-Contact Thermometer with Visible Field of View

  • Unique
    Technology

    Equipped with a ring laser pointer to improve work efficiency
  • First in
    Industry

    Measured temperature can be checked in a trend graph
  • Improved
    Performance

    High-accuracy measurement:
    ±1°C
    Response time:
    Achieves 50 ms or less
TI-S Series

Specifications

[Sensor Head]
Model TI-S30
Measurement range*1 −40 to +500°C
Display range*1 -50 to +510°C
Field of view φ30 mm at 500 mm
Optics Silicon lens
Sensing element/
spectral response
Thermopile 8 to 14 μm
Response time
(operating mode)
High speed, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s,
20 s
* Output response 90%*2
Accuracy*3 −40 to 0°C: ±1.5°C
+1 to +200°C: ±1°C
+201 to +500°C: ±0.5% of reading value
Repeatability ±0.5°C (when operating mode is 100 ms)
Temperature drift Within ±0.25°C/°C
Emissivity adjustment 0.100 to 1.200
Supply voltage 5 VDC (Supplied from controller)
Current consumption 30 mA or less/5 VDC
Connection type Pigtail cable 3 m
Minimum bending radius R 31 mm
Total cable length Max. 13 m (pigtail cable 3 m + extension cable 10 m)
Laser pointer Medium Red semiconductor laser
Wavelength 663 nm
Maximum output 1 mW
Laser class (JIS/IEC/FDA)*4 CLASS 2
Environmental
resistance
Degree of protection IP67 (IEC 60529)
Ambient temperature 0 to +80°C (up to +70°C during laser emission)
Ambient humidity 35 to 85% RH (no condensation)
Storage temperature −20 to +80°C
Vibration resistance 10 to 55 Hz Double amplitude 1.5 mm 2 hours in each of the X,Y and Z directions
Shock resistance 500 m/s2 (Approx. 50 G ) 3 times in each of the X,Y and Z directions
Applicable
regulations
EMC EMC Directive (2014/30/EU)
UK EMC (The Electromagnetic Compatibility Regulations 2016)
FCC Part 15 subpart B
Environment RoHS Directive (2011/65/EU), China RoHS (MIIT Order No. 32)
UK RoHS (The Restriction of the Use of Certain Hazardous
Substances in Electrical and Electronic Equipment Regulations 2012)
Safety FDA Regulation (21 CFR 1040.10 and 1040.11*5)
Applicable standards EN/IEC 61326-1
Material Case: Aluminum, Front plate: Stainless steel
Weight Approx. 180 g (including connector cable)

*1: If the measured temperature is below -50°C (lower limit display temperature), the displayed temperature is -50°C.
If the measured temperature is above 510°C (upper limit display temperature), the displayed temperature is 510°C.

*2: The response time is the time it takes for the output change to reach 90%.

*3: Measurement conditions: Emissivity; 1.000, Ambient temperature; 23 ±5°C, Size of the measurement target; sufficiently larger than the field of view.

*4: In accordance with the FDA provisions of Laser Notice No. 56, the laser is classified per the IEC 60825-1:2014 standard.

*5: Excluding differences per Laser Notice No. 56.

* To convert temperature values such as measurement temperature range and accuracy to Fahrenheit temperature, use (Fahrenheit temperature = Celsius temperature x 1.8 + 32).

* To convert relative values such as repeatability and temperature drift to Fahrenheit temperatures, use 1°C = 1.8°F.

● Specifications are subject to change without prior notice.

[Controller]
Model TI-SC (E)
Rating Supply voltage 24 VDC (when using a general-purpose I/O cable)
18 to 30 VDC (when using an IO-Link cable)
Current
consumption
180 mA (when using a general-purpose I/O cable)*1
50 mA (when using an IO-Link cable)
Display resolution 0.1°C/°F
Temperature unit Celsius “°C”/Fahrenheit “°F”
Measurement mode Normal/Sample hold/Peak hold/Valley hold/Edge detection
Response time
(operating mode)
High speed, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s
Output response 90%*2
Analog output/
IO-Link update time
High speed: 2.5 ms
50 ms to 2 s: 5.0 ms
5 s to 20 s: 100 ms
Analog
output
Resolution 10,801 steps
Accuracy Voltage ±0.2% of F.S. (at ambient temperature of 25°C)
Temperature coefficient (typical): ±22 ppm/°C (±0.0022%/°C)
Current ±0.2% of F.S. (at ambient temperature of 25°C)
Temperature coefficient (typical): ±4 ppm/°C (±0.0004%/°C)
Indicator Display 1.8-inch TFT LCD Display language: English, Simplified Chinese, Japanese
Power indicator When power is ON: lights in green, IO-Link communication: blinks in green
Output
indicator
Normal measurement alarm output is OFF: lights green
Normal measurement alarm output is ON: lights red
When minor warning occurs: blinks green
When major warning occurs: blinks orange
When error occurs: blinks red
Interface Alarm output NPN/PNP open collector (selectable by setting)
1 output: Max. 100 mA, 2 outputs: Max. 100 mA
Residual voltage NPN: 1.6 V or less, PNP: 3.4 V or less
Output mode N.O./N.C.
External input Laser off, Hold, Trigger
Analog
output
Current 4 to 20 mA load impedance: 150 to 500 ohm
Voltage 0 to 10 V output impedance: 200 ohm or less
Timer mode One shot/delay (ON delay, OFF delay)
One shot: 0.01 to 10.00 sec, Delay: 0.00 to 10.00 sec
IO-Link Revision 1.1.3
Baud rate COM 3 (230.4 kbps)
Number of process
input data bytes
4 bytes
Number of process
output data bytes
1 byte
Minimum cycle time 0.5 ms
Data storage class Data Storage Class 1: automatic DS
Connection
type
General-purpose
I/O cable
3 m cable 9 wires, Minimum bending radius: R 42 mm
IO-Link
cable
Open-end 3 m cable 3 wires, Minimum bending radius: R 42 mm
M12 4-pin
connector
0.3 m cable, Minimum bending radius: R 42 mm
Environmental
resistance
Degree of protection IP40 (IEC 60529)
Ambient temperature 0 to +50°C
Ambient humidity 35 to 85% RH (no condensation)
Storage temperature −20 to +70°C
Vibration resistance 10 to 55 Hz Double amplitude 1.5 mm 2 hours in each X, Y, Z direction
Shock resistance 500 m/s2 (Approx. 50 G ) 3 times in each of the X,Y and Z directions
Applicable
regulations
EMC EMC Directive (2014/30/EU)
UK EMC (The Electromagnetic Compatibility Regulations 2016)
FCC Part 15 subpart B
Environment RoHS Directive (2011/65/EU)
UK RoHS
(The Restriction of the Use of Certain Hazardous Substances
in Electrical and Electronic Equipment Regulations 2012)
China RoHS (MIIT Order No. 32)
Applicable standards EN/IEC 61326-1
Material Case: ABS
Weight Approx. 80 g

*1: Includes alarm output load current and analog output current.

*2: The response time is the time it takes for the output change to reach 90%.

* To convert temperature values such as measurement temperature range and accuracy to Fahrenheit temperature, use (Fahrenheit temperature = Celsius temperature x 1.8 + 32).

* To convert relative values such as repeatability and temperature drift to Fahrenheit temperatures, use 1°C = 1.8°F.

● Specifications are subject to change without prior notice.

Field of view(Unit: mm)


● The field of view is for a measurement diameter with an optical response (energy) of 90%.
● Size of the measurement target must be sufficiently larger than the field of view shown above.